BitScope DWG100

DSP Waveform Generator (Option for BS100)
DSP Waveform Generator
Next Item Prev Item Product Summary
Order Code DWG100
Description BitGen integrated waveform and timing generator solution for USB BitScope 100. Pre-installed if purchased with BS100U
Comment Option for BS100.
Price
Quantity USD (US Dollar)
1+ $ 84.95

Product Information

DWG100 (BitGen) is a DSP based waveform and timing generator upgrade for BS100U.

It mounts inside the BitScope and comes with installation instructions and new software. It is a standard feature of BS120U, you don't need to purchase this upgrade if you have BS120U.

Waveform Generators Specification
Generator Types [1] VSR, DDS, CLK & PRNG
Frequency (max) [2] 1MHz (DDS, PRNG), 1.5MHz (VSR) or 5MHz (CLK)
Frequency (min) [2] 1mHz (DDS), 1Hz (VSR, PRNG) or 1kHz (CLK)
Resolution 6 digits or 1mHz (DDS)
Accuracy [3] ±30ppm, 18° to 30° (typical)
Temperature Coefficient [4]≈1ppm/°C
Built-in Wave-Functions Sine, Ramp, Step, Exponential, Noise
Loadable Wavetable Size 2 to 1024 samples
Trigger/Burst/Clock [5] Specification
Operating Modes Burst, Gated, Sweep, Clock In/Out
Burst Duration [6] 1uS to >1Hr
Post-Trigger Delay [6] 500nS to >1Hr
Inter-burst Delay [6] 500nS to >1Hr
Burst Repeat Count 2^32 (max)
WavePort Specification
Output Level 10Vpp to 40mVpp
Output Offset ±2.5Vpp
Output Impedance 50Ω
Output Voltage Tolerance ±12V (max)
Waveform Resolution (D/A) 8 bits (multiplying)
TimePort Specification
Output Type 3-state, 100k to GND
Logic Levels 5V CMOS In/Out
Output Drive Active (low ≈0.6V, high ≈4.4V)
Output Impedance 100Ω (50mA max)
Trigger Latency 200nS (typical)
External Clock (max) 2MHz
Minimum Pulse Width 100nS
[1] Supports 4 different generator types: VSR = Variable Sample Rate, DDS = Direct Digital Synthesis, CLK = Clock Generator, PRNG = Psuedo Random Number Generator.
[2] Minimum and maximum frequency depends on generator type as indicated.
[3] Uncalibrated frequency accuracy is specified (as defined by the BS100U reference crystal). May be calibrated to better than this if precision frequency reference is available.
[4] Temperature Coefficient is that of the BS100U reference crystal.
[5] Trigger, Burst and Clock modes may be configured to accept or generate a trigger or clock signal via the TimePort.
[6] The maximum duration or delay depends on the sample rate. The maximum duration in all cases exceeds 1 hour but can extend far beyond this limit at lower sample rates.